From: Pressure sensor calibrations of acoustic telemetry transmitters

Fig. 3

Boxplots showing a slopes, b intercepts, and c R 2 values of sensor values versus actual depth values of sensor values versus actual depth of ten acoustic transmitters when calibrated in a pressurized chamber at Carleton University (Ottawa, ON) at three different temperatures (cold = 9 °C, cool = 20 °C, warm = 34 °C). Black lines within boxes represent medians, end of boxes represent the 25 and 75th percentiles, and whiskers represent the 10 and 90th percentiles. Solid circles indicate outliers outside the 10 and 90th percentiles. Bars not sharing the same letter are significantly different at p < 0.05 (Tukey HSD test)