Skip to main content

Table 1 Artificial ECG signal fed at an accurate frequency generated by signal generator to the tag

From: Validation of ECG-derived heart rate recordings in Atlantic cod (Gadus morhua L.) with an implantable data logging system

Frequency–bpm

Manual calculation [bpm]

Algorithmic calculation [bpm]

0.5 Hz–30 bpm

30

29

1 HZ–60 bpm

60

60

2 Hz–120 bpm

120

120

3 Hz–180 bpm

180

181

4 Hz–240 bpm

240

240

5 Hz–300 bpm

300

300

  1. Manual calculation from the ECG, recorded with the logger and the Algorithmic calculated values presented in bpm. Sampling frequency of the ECG is 100 Hz